Introduction of IC6600 Ion Chromatography Machine
The IC6600 multifunctional ion chromatograph adopts a brand-new modular design, offering flexible configuration and powerful performance. The system supports time-sharing injection, simultaneous injection, and online enrichment autosamplers, enabling automated operations ranging from simple sample introduction to complex sample processing. The "water-only" eluent generator can produce the required eluent online, achieving gradient elution under isocratic pump conditions. This reduces costs, significantly improves the reproducibility of experimental results while minimizing labor requirements, and lowers the risk of contamination.
Technical Parameters of IC6600 Ion Chromatograph:
| Product Name | Multifunctional Ion Chromatography |
| Model | IC6600 |
| Working power | 220V, 50Hz |
| Ambient temperature | 10-40°C |
| Relative humidity | 20-85% |
| Qualitative repeatability | ≤0.1% |
| Quantitative repeatability | ≤0.5% |
| Pump flow rate range | 0.001~12.000mL/min |
| Pressure resistance of pump head | 0-40MPa |
| Flow setting error | <0.1% |
| Flow stability | <0.1% |
| Pressure display accuracy | 0.001MPa |
| Pressure plusation | <0.25% |
| Conductivity detector | |
| Baseline noise | ≤0.03%FS |
| Baseline drift | ≤0.1%FS |
| Mini detectable concentration (Cl): | ≤0.0001μg/mL |
| Mini detectable concentration (Li): | ≤0.00005μg/mL |
| Mini detectable concentration (NO3): | ≤0.001μg/mL |
| Mini detectable concentration (Na+): | ≤0.001μg/mL |
| Mini detectable concentration (Ca2+): | ≤0.005μg/mL |
| Conductivity cell volume | ≤0.4μL |
| Conductivity detection range | 0-45000μS/cm |
| Detector resolution | 0.00238nS/cm |
| Conductivity cell temperature setting error | ±0.01°C |
| Conductivity cell temperature display resolution | 0.001°C |
| Conductivity cell temperature control range | Detector temperature zone temperature +5°C-65°C |
| Conductivity cell temperature control stability | 0.05°C |
| Conductivity cell temperature setting value error | 0.1°C/h |
| Ampere detector | |
| Type | Thin-layer structure, automatic range |
| Driving voltage range | -2.0V-2.0V |
| Drive voltage resolution | 0.001V |
| Output signal range |
Integral mode: 50pc-200uC DC mode: 1pA-74μA |
| Noise |
Integration mode<30pC DC mode<5pA, DC mode static noise<1pA |
| Cell material | Titanium |
| Pool volume | <0.2μL |
| Maximum pool pressure | 0.7MPa |
| Minimum detectable concerntration | I-0.001μg/mL |
| Minimum detectable concerntration fructose | 0.01μg/mL |
| UV Detector | |
| Wavelength indication error | 0.1nm |
| Wavelength repeatability | 0.1nm |
| Mini detectable concentration | BrO3-0.0005μg/mL |
| Mini detectable concentration | CrO42-0.001μg/mL |
| Qualitative repeatability | ≤0.05% |
| Quantitative repeatability | ≤0.3% |
| Suppressor | |
| Dead volume | 50μL |
| Inhibition capacity | 170ueq/min |
| Pressure resistance | 3MPa |
| Column oven | |
| Temperature range | Room temperature +5°C~70°C |
| Temperature control stability | 0.1°C/h |
| Error of temperature setting value | 0.05°C |
| Autosampler | |
| Number of samples | ≥120 (1.5ml sample bottle) |
| Temperature | 4~30°C |
| Linear | >0.999 |
| Cross-contamination | <0.01% |