Place of Origin:
Anhui, China
Brand Name:
Wayeal
Certification:
ISO, CE
Model Number:
IC6200
Wayeal Laboratory Analytical Ion Chromatography Instrument With Autosampler
Introduction of Ion Chromatography Instrument
Ion chromatography instrument is a type of high performance liquid chromatograph, so it is also known as high performance ion chromatography (HPIC) or modern ion chromatography. IC systems can analyze particles such as anions, cations, organic salts, and proteins. They are used in environmental, manufacturing, food, pharmaceutical, and chemical industries.
Parameters of Ion Chromatography Instrument
Wayeal IC6200 Integrated Ion Chromatography Instrument for Cations or Anions Analysis | |
System Parameter | |
Qualitative Repeatability | ≤0.5% |
Quantitative Repeatability | ≤1% |
Pump System | |
Flow Rate | (0.001~10.000)mL/min |
Pump Head Pressure Resistance | (0 ~ 35) MPa |
Flow Rate Setting Inaccuracy | <0.1% |
Flow Stability | <0.1% |
Pressure Display Accuracy | 0.01Mpa |
Pressure Pulsation | <0.5% |
Conductivity Detector | |
Baseline Noise | ≤0.1%FS |
Baseline Drift | ≤0.5%FS |
Minimum Detectable Concentration (Cl) | ≤0.0002μg/mL |
Minimum Detection Concentration (Li) | ≤0.0001μg/mL |
Conductivity Cell Volume | ≤ 0.6μL |
Conductivity Detection Range | 0-15000μS/cm |
Detector Resolution | 0.00238nS/cm |
Conductivity Cell Temperature Setting Error | ±0.01°C |
Resolution of Conductivity Cell Temperature Display | 0.001°C |
Suppressor | |
Constant Current Source Range | 0-500mA in 0.1mA increments |
Column Oven | |
Temperature Range | Room temperature +5°C~85°C |
Control Temperature Stability | ≤ 0.1°C |
Autosampler | |
Number of Sample Vials | ≥ 120 (1.5mL standard sample bottle) |
Repeatability |
Full loop injection < 0.3% RSD, Partial injection < 0.5% RSD, Non-destructive injection < 1 % RSD (injection volume > 5uL) |
Syringe Size | Standard: 500μL; optional; 250μL, 1000μL, 2500μL |
Loop Size | Standard: 100μL; optional: 20μL, 50μL, 200μL |
Linearity | >0.999 |
Cross-contamination | <0.01% |
Eluent Generator | |
Flow Rate Range | 0.001 to 3.000mL/min |
Eluent Concentration Range | KOH, MSA-0.1 to 100mM |
Concentration Accuracy | 0.1mM |
Original Concentration of Electrolyte | 25% KOH/MSA |
Volume of Electrolyte | 1000mL |
Maximum Working Pressure | 21MPa (3000psi) |
Gradient Accuracy | 1.5% |
Performance of Ion Chromatography Instrument
1. Highly accurate and efficient temperature-controlled conductivity cell ensures the stability and high sensitivity of the assay, as well as the stability of the instrument under any laboratory conditions.
2. Simple workstation interface, easy to use processes, and full of detailed management.
3. Automatic and intelligent functions such as bottle shortage alarm, liquid leakage alarm and door opening/closing sensor light to ensure reliable operation of the equipment.
4. The full system thermostat and temperature compensation function allows you to carry out your analytical work without the distraction of controlling the temperature conditions in the laboratory and without the need for air conditioning.
5.Tandem dual plunger mode with electronic pulsation suppression technology provides high precision, low pulsation and low drift drenching solutions.
Industry Application
Environment, food, medicine, agriculture, grain and oil, and chemical industry
Send your inquiry directly to us