Place of Origin:
Anhui, China
Brand Name:
Wayeal
Certification:
CE
Model Number:
IC6300
Product Description:
Ion chromatography (or ion-exchange chromatography) is a form of chromatography that separates ions and ionizable polar molecules based on their affinity to the ion exchanger. It works on almost any kind of charged molecule—including small inorganic anions, large proteins, small nucleotides, and amino acids. However, ion chromatography must be done in conditions that are one pH unit away from the isoelectric point of a protein.
42PMa Modular Design Ion Chromatography Instrument with Ampere Detector |
|
---|---|
Product Name | Ion Chromatography Instrument |
Working Conditions | |
Working Power | 220V, 50Hz |
System Parameters | |
Qualitative Repeatability | ≤1% |
Quantitative Repeatability | ≤1% |
Pump System | |
Flow Rate | 0.001~15.000mL/min |
Pressure Range | 0~42MPa |
Flow Rate Setting Error | <0.2% |
Flow Rate Stability | <0.2% |
Eluent | |
Flow Rate Range | 0.001~3.000ml/min |
Eluent Concentration Range | KOH/MSA 0.1~100.0mM |
Accuracy of Concentration | 0.1mM |
Max Working Pressure | 21MPa (3000psi) |
Autosampler (AS2800 series) | |
Quantitative Repeatability | Full loop injection <0.5%RSD |
Linearity | >0.9995 |
Injection Positions | 48 Positions |
Working Temperature | 10~40°C |
Sample Bottle | 5mL/100mL |
Injection Volume | 10uL~400uL |
Autosampler (AS3110 Series) | |
Injection Positions | 120/64/92 positions for optional |
Cooling Temperature | 4~30°C |
Injector | Standard 500ul, options 250ul, 100ul, 2500ul |
Column Oven | |
Temperature Range | Room temperature, +5℃~70℃ |
Temperature Control Stability | ≤0.1°C |
Temperature Setting Error | 0.05°C |
Column Capacity | 250mm and 150mm |
Suppressor | |
Dead Volume | 50uL |
Suppression Capacity | 170ueq/min |
Pressure Resistance | 3MPa |
Conductivity Detector | |
Baseline Noise | ≤0.0005μS/cm |
Baseline Drift | ≤0.0005μS·cm-1/30min |
Min Detection Concentration(Cl) | ≤0.0005ug/mL |
Min Detection Concentration (Li) | ≤0.0002ug/mL |
Cell Volume | ≤0.4μL |
Detector Resolution | 0.00238nS/cm |
Temperature Setting Error | ±0.01°C |
Ampere Detector | |
Drive Voltage Range | -2.048V---2.048V |
Drive Voltage Resolution | 0.001V |
Single Input Range | Integral mode, 50pc---200uC |
DC mode | 1pA---74uA |
Cell Volume | <0.2uL |
Min Detection Concentration | I: 0.001μg/mL |
Application of Ion Chromatography Instrument
The Ion Chromatography Instrument is suitable for use in a variety of applications, including:
Environmental Testing: Analysis of water, soil, and air samples for pollutants and contaminants.
Pharmaceuticals: Determination of ionic impurities in drugs and pharmaceutical products.
Food and Beverage: Analysis of ionic species in food and beverage products for quality control and safety.
Chemical Analysis: Detection and quantification of ionic species in chemical compounds and mixtures.
Self-created eluent liquid purification system, effectively removes trace impurity ions and provides ultra-pure eluent liquid system.
Multiple temperature control, always ensure system uniformity and stability. Separation, detection and multiple temperature control modes provide higher temperature stability.
The highest pressure resistant all-PEEK ion chromatography pump with a pressure resistance of up to 42Mpa;
Send your inquiry directly to us