Place of Origin:
Anhui, China
Brand Name:
Wayeal
Certification:
ISO, CE
Model Number:
CEMS1900
Document:
FTIR Real-Time Emission Monitoring System for Industrial Pollutants
Product Introduction
The FTIR Real-Time Emission Monitoring System employs Fourier Transform Infrared (FTIR) absorption technology to measure gaseous pollutants in flue gases from sources such as municipal solid waste incineration, hazardous waste incineration, and solid waste incineration. It consists of a pretreatment unit, an analysis unit, and a data acquisition and processing unit, capable of continuously monitoring wet basis, dry basis, and converted values of parameters including SO₂, NO, NO₂, HF, HCl, CO, CO₂, H₂O, and NH₃. Additionally, based on multiple relevant parameters such as flue gas temperature, pressure, and flow rate, it calculates emission rates and total emission amounts.
Product Parameter
|
Gases |
Parameter |
|
Range (SO2) |
(0~75)mg/m3,(0~2000)mg/m3, accept customized |
|
Range (NO) |
(0~100)mg/m3,(0~2000)mg/m3, accept customized |
|
Range (NO2) |
(0~100)mg/m3,(0~1000)mg/m3, accept customized |
|
Range (NH3) |
(0~5)mg/m3,(0~50)mg/m3, accept customized |
|
Range (HCL) |
(0~100)mg/m3,(0-150)mg/m3, accept customized |
|
Range (HF) |
(0~3)mg/m3,(0~50)mg/m3, accept customized |
|
Range (CO) |
(0~100)mg/m3,(0~1000)mg/m3, accept customized |
|
Range (H2O) |
(0~40)% |
|
Range (CO2) |
(0~25)% |
|
Range (O2) |
(0~25)% |
|
Zero Drift |
≤+2%F.S. |
|
Range Drift |
≤+2%F.S. |
|
Linear Error |
≤+2%F.S. |
|
Response Time |
≤200s, HCL, HF≤400s(50m heat tracing pipeline) |
System Characteristics
Multiple Measurement Substances
Simultaneous measurement of SO2, NOx, (NO+NO2), NH3, HCI, HF, CO, CO2, H2O and other components is possible, and other components can be added as required.
Anti-interference
1. Algorithm optimization, which can effectively eliminate the interference of components such as H2O when choose a suitable wavelength range.
2. With CO2 scrubber and dryer, the instrument air can be converted into purge gas and zero gas suitable for Fourier infrared spectrometer conditions, with automatic zero calibration function.
Industry Application
Steel plants, cement plants, chemical plants, etc.