Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

2026-09-11

Latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography
Case Detail

Organochlorine pesticides are a class of synthetic persistent organic pollutants that have been extensively used as broad-spectrum insecticides. These pesticides are extremely stable in chemical properties and difficult to degrade, and can persist in the environment for long periods. They can enter ambient air through atmospheric transport, dry and wet deposition, and other processes, posing a long-term threat to ecosystems and human health.

This article refers to the standard "Ambient Air Determination of Organochlorine Pesticides Gas Chromatography" (HJ 901-2017) and employs the Wayeal GC6100 gas chromatograph equipped with an ECD detector for the determination of 23 organochlorine pesticides in ambient air.

Keywords: Organochlorine pesticides; Gas chromatography; ECD detector; Ambient air.

1. Instrument Configuration

1.1 Instrument Configuration

Table 1 Configuration List of Gas Chromatograph

No.

Modular

Qty

1

GC6100 Gas Chromatograph

1

2

ECD Detector

1

3

ALS6100 Autosampler

1

1.2 Experimental Materials and Auxiliary Equipment

23 organochlorine standard stock solution (1000μg/mL): commercially purchased certified standard solution, stored sealed and protected from light under refrigeration below 4 °C.

Surrogate standard stock solution (1000μg/mL): commercially purchased certified standard solution, containing a mixture of 2,4,5,6-tetrachloro-m-xylene (TCX) and decachlorobiphenyl (DCBP), stored sealed and protected from light under refrigeration below 4 °C.

Internal standard stock solution (5000μg/mL): commercially purchased certified standard solution, containing 1-bromo-2-nitrobenzene (BNB), stored sealed and protected from light under refrigeration below 4 °C.

n-Hexane: pesticide residue grade

Carrier gas: high-purity nitrogen.

1.3 Test Conditions

Chromatographic column: quartz capillary column, stationary phase of 14% cyanopropylphenyl86% methylpolysiloxane, dimensions 30 m × 0.32mm × 0.25μm.

1.4 Solution Preparation

1.4.1 Internal Standard Intermediate Solution (100μg/mL):

Pipette 1.0mL of the internal standard stock solution (5000μg/mL) into a 50mL volumetric flask, dilute to volume with n-hexane, and mix thoroughly.

1.4.2 Internal Standard Working Solution (10μg/mL)

Pipette 1.0mL of the internal standard intermediate solution (100μg/mL) into a 10mL volumetric flask, dilute to volume with n-hexane, and mix thoroughly.

1.4.3 Surrogate Intermediate Solution (50μg/mL):

Pipette 0.5mL of the surrogate standard stock solution (1000μg/mL) into a 10mL volumetric flask, dilute to volume with n-hexane, and mix thoroughly.

1.4.4 Surrogate Working Solution (1μg/mL):

Pipette 1.0 mL of the surrogate intermediate solution (50 μg/mL) into a 50 mL volumetric flask, dilute to volume with n-hexane, and mix thoroughly.

1.4.5 23-Component Organochlorine Standard Intermediate Solution (40μg/mL):

Pipette 1.0mL of the 23-component organochlorine standard stock solution (1000μg/mL) into a 25mL volumetric flask, dilute to volume with n-hexane, and mix thoroughly.

1.4.6 23-Component Organochlorine Standard Working Solution (1μg/mL):

Pipette 250μL of the standard intermediate solution (40μg/mL) and 200μL of the surrogate intermediate solution (50μg/mL) into a 10mL volumetric flask, dilute to volume with n-hexane, and mix thoroughly.

1.4.7 Linear Calibration Standard Working Solutions

Pipette appropriate amounts of the 23-component organochlorine standard working solution (1μg/mL) and dilute with n-hexane to prepare a calibration standard series at concentrations of 20μg/L, 50μg/L, 100μg/L, 200μg/L, and 300μg/L. Quantification is performed by the internal standard method, with 10.0μL of the internal standard working solution (10μg/mL) added to every 1.0mL of standard solution.

2. Result and Discussion

2.1 Qualitative Test of Standard Samples

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

Fig 1 Chromatogram of the 23-organochlorine Pesticide Standard Solution (300μg/L)

(Elution order: 1: BNB (internal standard), 2: TCX (surrogate), 3: hexachlorobenzene, 4: α-HCH, 5: γ-HCH, 6: heptachlor, 7: aldrin, 8: β-HCH, 9: δ-HCH, 10: heptachlor epoxide, 11: endosulfan I, 12: γ-chlordane, 13: α-chlordane, 14: 4,4'-DDE, 15: dieldrin, 16: endrin, 17: 2,4'-DDT, 18: 2,4'-DDD, 19: endosulfan II, 20: 4,4'-DDT, 21: endrin aldehyde, 22: mirex, 23: methoxychlor, 24: endosulfan sulfate, 25: endrin ketone, 26: decachlorobiphenyl (surrogate).)

Table 2 Chromatographic Parameters of the 23-Organochlorine Pesticide Standard Solution (300μg/L)

Compound

Retention Time (min)

Peak Area

Theoretical Plate Number

Resolution

BNB (Internal Standard)

11.241

92.289

244468

37.351

TCX (Surrogate)

15.233

275.604

246274

21.733

Hexachlorobenzene

18.066

355.274

274493

20.273

α-HCH

20.852

357.157

370716

17.819

γ-HCH

23.353

346.851

422627

7.218

Heptachlor

24.417

334.795

417131

10.064

Aldrin

25.951

341.574

456981

14.983

β-HCH

28.189

174.309

603131

8.681

δ-HCH

29.435

316.470

689426

5.782

Heptachlor Epoxide

30.290

293.621

617482

9.330

Endosulfan I

31.759

284.150

625489

2.635

γ-Chlordane

32.181

311.083

648301

2.839

α-Chlordane

32.640

301.553

639967

3.408

4,4'-DDE

33.177

296.160

765089

4.977

Dieldrin

33.954

310.288

715965

7.033

Endrin

35.118

281.535

678092

4.151

2,4'-DDT

35.810

190.808

775595

12.785

4,4'-DDD

37.811

245.570

1011677

1.348

Endosulfan II

38.022

274.936

863408

4.239

4,4'-DDT

38.707

203.469

943775

7.173

Endrin Aldehyde

40.456

173.135

242878

2.190

Mirex

41.016

239.629

806372

7.207

Methoxychlor

42.281

254.181

1009022

1.070

Endosulfan Sulfate

42.459

126.024

1068505

11.123

Endrin Ketone

44.336

237.775

1048460

28.722

Decachlorobiphenyl (Surrogate)

49.407

281.111

1206320

N/A

Note: As can be seen from the above chromatogram, the resolution between the chromatographic peaks of all substances in the standard solution is greater than 1.0, which meets the requirements of experimental analysis.

2.2 Linearity

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

Fig 2 Calibration Curves and Correlation Coefficients of the 23 Organochlorine Pesticides

Note: The linearity of each component in the 23-organochlorine pesticide standard solution is good, with linear correlation coefficients > 0.999, which is higher than the standard requirement (linear correlation coefficient0.995) and meets the requirements of experimental analysis.

2.3 Precision

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

Fig 3 Chromatogram of the 23-organochlorine Pesticide Standard Solution (low concentration, 20μg/L)

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

Fig 4 Chromatogram of the 23-organochlorine Pesticide Standard Solution (medium concentration, 100μg/L)

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

Fig 5 Chromatogram of the 23-organochlorine Pesticide Standard Solution (high concentration, 300μg/L)

Table 3 Chromatographic Parameters for Precision

Low Concentration (20μg/L)

Medium Concentration (100μg/L)

High Concentration (300μg/L)

Component

RSD%

Component

RSD%

Component

RSD%

BNB (Internal Standard)

1.979

BNB (Internal Standard)

1.264

BNB (Internal Standard)

1.238

TCX (Surrogate)

1.400

TCX (Surrogate)

1.173

TCX (Surrogate)

1.334

Hexachlorobenzene

2.207

Hexachlorobenzene

1.255

Hexachlorobenzene

0.850

α-HCH

1.298

α-HCH

1.072

α-HCH

1.379

γ-HCH

1.828

γ-HCH

0.974

γ-HCH

1.468

Heptachlor

2.057

Heptachlor

1.088

Heptachlor

1.437

Aldrin

1.601

Aldrin

0.891

Aldrin

1.233

β-HCH

1.529

β-HCH

0.656

β-HCH

0.575

δ-HCH

1.359

δ-HCH

1.257

δ-HCH

1.354

Heptachlor Epoxide

1.385

Heptachlor Epoxide

1.181

Heptachlor Epoxide

0.960

Endosulfan I

1.784

Endosulfan I

0.988

Endosulfan I

1.439

γ-Chlordane

2.021

γ-Chlordane

1.005

γ-Chlordane

1.480

α-Chlordane

1.961

α-Chlordane

1.029

α-Chlordane

0.356

4,4'-DDE

1.313

4,4'-DDE

0.977

4,4'-DDE

1.508

Dieldrin

1.314

Dieldrin

1.041

Dieldrin

1.429

Endrin

1.827

Endrin

1.219

Endrin

1.488

2,4'-DDT

2.000

2,4'-DDT

1.165

2,4'-DDT

0.748

4,4'-DDD

0.786

4,4'-DDD

1.205

4,4'-DDD

1.575

Endosulfan II

1.740

Endosulfan II

1.036

Endosulfan II

1.559

4,4'-DDT

1.259

4,4'-DDT

1.571

4,4'-DDT

1.120

Endrin Aldehyde

2.036

Endrin Aldehyde

1.375

Endrin Aldehyde

0.877

Mirex

1.769

Mirex

1.225

Mirex

0.961

Methoxychlor

1.216

Methoxychlor

1.226

Methoxychlor

0.937

Endosulfan Sulfate

1.041

Endosulfan Sulfate

1.067

Endosulfan Sulfate

0.806

Endrin Ketone

1.950

Endrin Ketone

1.153

Endrin Ketone

1.421

Decachlorobiphenyl (Surrogate)

1.953

Decachlorobiphenyl (Surrogate)

1.353

Decachlorobiphenyl (Surrogate)

1.467

Note: The 23-organochlorine pesticide standard solutions at the lowest, middle, and highest concentration points of the calibration curve were each determined in 5 replicate measurements. Their relative standard deviations (RSDs) were 0.82.2%, 0.71.6%, and 0.41.6%, respectively. The RSDs of the chromatographic peaks for all substances meet the standard requirements.

2.4 Limit of Detection

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

Fig 6 Chromatogram of the Detection Limit Solution (10μg/L)

Table 4 Method Detection Limits and Lower Limits of Determination for Each Component of the 23 Organochlorine Pesticides

latest company case about Determination of 23 Organochlorine Pesticides in Ambient Air by Wayeal Gas Chromatography

The 23-organochlorine pesticide standard solution (10μg/L) was determined in 8 replicate measurements. Calculations showed that, with a sampling volume of 350m³ and a concentration volume of 1.0mL, the method detection limits for the 23 organochlorine pesticides were in the range of 0.0003ng/m³ to 0.0025ng/m³, and the lower limits of determination were in the range of 0.0012ng/m³ to 0.010ng/m³, which comply with the standard requirements.

3. Conclusion

This method uses the Wayee GC6100 gas chromatograph equipped with an ECD detector for the determination of 23 organochlorine pesticides in ambient air. The experimental results show that the resolution between the chromatographic peaks of all organochlorine pesticides is greater than 1.0, which meets the requirements of experimental analysis. The linearity of each component in the 23-organochlorine pesticide standard solution is good, with linear correlation coefficients > 0.999, which is higher than the standard requirement (linear correlation coefficient 0.995). The precision test results, as well as the method detection limits and lower limits of determination, all meet the standard requirements. Therefore, this method can be adopted for the determination of 23 organochlorine pesticides in ambient air.